IRECTION 5535208-100, REV. 2 LOGIQ E9 SERVICE MANUAL
4 - 38 Section 4-3 - Functional checks
4-3-8 Tissue Velocity Imaging (TVI) Checks
4-3-8-1 Introduction
TVI calculates and color codes the velocities in tissue. The tissue velocity information is acquired by
sampling of tissue Doppler velocity values at discrete points. The information is stored in a combined
format with grey scale imaging during one or several cardiac cycles with high temporal resolution.
4-3-8-2 Preparations
1.) Connect one of the probes, to the scanner’s left-most probe connector.
-See: Section 3-6-6 "Connecting Probes" on page 3-20 for info about connecting the
probesSection 9-18 "Probes" on page 9-104
2.) Turn ON the scanner
The 2D Mode window is displayed (default mode).
3.) If needed, adjust the Display’s Brightness and Contrast setting.
4.) Press TVI.
5.) Use the trackball (assigned function: Pos) to position the ROI frame over the area to be examined.
6.) Press Select. The instruction Size should be highlighted in the trackball status bar.
NOTE: If the trackball control pointer is selected, press trackball key to be able to select between
Position and Size controls.
7.) Use the trackball to adjust the dimension of the ROI.
4-3-8-3 Adjust the TVI Controls
To reduce quantification noise (variance), the Nyquist limit should be as low as possible, without
creating aliasing. To reduce the Nyquist limit: Reduce the Scale value.
NOTE: The Scale value also affects the frame rate. There is a trade off between the frame rate and
quantification noise.
TVI provides velocity information only in the beam direction. The apical view typically provides the
best window since the beams are then approximately aligned to the longitudinal direction of the
myocardium (except near the apex). To obtain radial or circumferential tissue velocities, a
parasternal view must be used. However, from this window the beam cannot be aligned to the
muscle for all the parts of the ventricle.
NOTE: PW will be optimized for Tissue Velocities when activated from inside TVI.