GE
D
IRECTION 5535208-100, REV. 2 LOGIQ E9 SERVICE MANUAL
7 - 40 Section 7-5 - Common Service Desktop
7-5-8 Diagnostics for Service Diagnostics
7-9-19 "LOGIQ E9 Transmit and Receive Signal Path Diagnostics" on page 7-268. Also for 7-5-12-15
"Diagnostics Window Overview - R3.x.x and later" on page 7-96
7-5-8-1 Digital Receive
See: 7-5-3 "Diagnostics Window Overview" on page 7-22 for a description of the looping and list
features.
Symptoms:
Intermittent general problems (communication between DRX and GFI)
Image artifacts - pixilated/jagged edges
1.) Select the Digital Receive checkbox to run all the tests, or select the individual subtests to run only
the selected tests.
2.) Enter the desired Loop Count (numeric value of 4 digits or less).
3.) Click Execute to run the test(s).