IRECTION 5535208-100, REV. 2 LOGIQ E9 SERVICE MANUAL
7 - 46 Section 7-5 - Common Service Desktop
7-5-8-3 GFI
See: 7-5-3 "Diagnostics Window Overview" on page 7-22 for a description of the looping and list
features.
Symptoms:
TGC Problems
Doppler Audio Problems
Image Artifacts
Intermittent Instability
1.) Select the GFI checkbox to run all the tests, or select the individual subtests to run only the selected
tests.
2.) Enter the desired Loop Count (numeric value of 4 digits or less).
3.) Click Execute to run the test(s).