D
IRECTION 5535208-100, REV. 2 LOGIQ E9 SERVICE MANUAL
Chapter 7 Diagnostics/Troubleshooting 7 - 51
7-5-8-5 Noise
See: 7-5-3 "Diagnostics Window Overview" on page 7-22 for a description of the looping and list
features.
Symptoms:
Image artifacts
•Image noise
Poor image quality
Missing image channels
1.) Select the Noise checkbox or the individual subtest to run the test.
2.) Enter the desired Loop Count (numeric value of 4 digits or less).
3.) Click Execute to run the test.