IRECTION 5535208-100, REV. 2 LOGIQ E9 SERVICE MANUAL
Chapter 7 Diagnostics/Troubleshooting 7 - 63
7-5-8-10 I/O Board Tests
See: 7-5-3 "Diagnostics Window Overview" on page 7-22 for a description of the looping and list
features.
Symptoms:
Intermittent Problems
Popup messages warning of system voltage problems - I/O Board Pwr Supply Test
System comes up in simulator mode - run I/O Board GFI Voltage Test
Unable to run diagnostics
No doppler audio/system audio/general audio sounds - run I/O Board Audio Test
No display on the monitor
1.) Select the I/O Board Tests checkbox to run all the subtests, or select the individual subtests to run
only the selected tests.
2.) Enter the desired Loop Count (numeric value of 4 digits or less).
3.) Click Execute to run the test(s).