IRECTION 5535208-100, REV. 2 LOGIQ E9 SERVICE MANUAL
7 - 76 Section 7-5 - Common Service Desktop
7-5-10-3 System Board
See: 7-5-3 "Diagnostics Window Overview" on page 7-22 for a description of this window’s features.
Runs tests on essential functions of the PC Mother Board. The test results are displayed in the Status
portion of the window.
Run Time: About 1 minute.
1.) Enter the desired Loop Count (numeric value of 4 digits or less).
2.) Click the Execute button. If there are any failures:
3.) Check the Motherboard Harness. Swap with a known good harness. If the problem continues:
4.) Replace the Motherboard. If the problem continues:
5.) Replace the BEP.