GE
D
IRECTION 5535208-100, REV. 2 LOGIQ E9 SERVICE MANUAL
Chapter 7 Diagnostics/Troubleshooting 7 - 81
7-5-10-8 Network Adapter Diagnostic Changes for R4
Network Adapter Test is split into 2 folders:
• Network Adapter - see: Table 7-124 on page 7-157 and Table 7-125 on page 7-158
• Network Adapter Reliability - see: Table 7-126 on page 7-159 and Table 7-127 on page 7-160
Each folder contains separate tests for Wired and Wireless Adapters.
7-5-10-9 Memory
See: 7-5-3 "Diagnostics Window Overview" on page 7-22 for a description of this window’s features.
Runs a test on essential functions of the Memory. The test results are displayed in the Status portion of
the window.
1.) Enter the desired Loop Count (numeric value of 4 digits or less).
2.) Click the Execute button.
3.) Replace the BEP if any tests fail.