IRECTION 5535208-100, REV. 2 LOGIQ E9 SERVICE MANUAL
7 - 90 Section 7-5 - Common Service Desktop
7-5-15 Patient I/O Tests
Under Service Diagnostics / Patient I/O.
This procedure provides a basic test of the ability of the system to start and stop the acquisition of data
from the Patient I/O Module to the Host via the GFI.
INPUT > The connection of ECG leads, PHONO (PCG) or AUX is optional.
OUTPUT > The results display information about the signal data acquired from the ECG input to the
Patient I/O Module. The information includes the minimum signal value, maximum signal value and
mean signal value.
If this test fails:
1.) Run GFI tests
If GFI is ok:
2.) Replace Patient I/O module
If this test passes and there are still issues with poor or no ECG trace, replace ECG leads.