IRECTION 5535208-100, REV. 2 LOGIQ E9 SERVICE MANUAL
Chapter 8 Replacement procedures 8 - 5
8-2-5 Tools needed for servicing the LOGIQ E9
In addition to the standard tools listed in 10-4-2 "“Standard” GE Tool Kit in the USA" on page 10-4, the
following tools (TORX bits or drivers) are needed to service the LOGIQ E9. If the torque is not indicated
with the procedure, hand tighten the screws/nuts.
Table 8-1 Tools used for servicing LOGIQ E9
TOOL TYPE TOOL SIZE
TORX
BIT # TX-10
BIT # TX-15
BIT # TX-20
BIT # TX-25
BIT # TX-30*
BIT # Tx-45*
FLAT BLADE
SCREWDRIVER
1/4 inch or 6 mm
PHILLIPS
SCREWDRIVER
PH1
PH2
PH3
HEX KEY
5 mm
8 mm
10 mm
OPEN END WRENCH
TORQUE WRENCH
5 mm
10 mm
13 mm
3/16 inch
Up to 81 Nm (59.7 lbf-ft)**
ADJUSTABLE
WRENCH
152 mm (6 inches) long
PRY BAR approximately 18 inches overall (457 mm)
* 90 degree “L“ is suggested. A full set of 90 degree “L“ Torx wrenches are
recommended.
** Heavy mechanical parts may need a specific torque. Each procedure will
indicate the torque needed.

Table of Contents

Other manuals for GE LOGIQ E9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ E9 and is the answer not in the manual?

GE LOGIQ E9 Specifications

General IconGeneral
TypeUltrasound System
3D/4D ImagingYes
Transducer Ports4 active ports
ConnectivityDICOM, USB, Ethernet
ApplicationsAbdominal, Vascular, Musculoskeletal, Small Parts, Urology, Pediatric, Cardiac, Obstetrics/Gynecology
TransducersLinear, Convex, Phased Array, Volume
Doppler ModesColor Doppler, Power Doppler, Pulsed Wave Doppler, Continuous Wave Doppler
ElastographyYes
Advanced FeaturesB-Flow, Volume Navigation
PortabilityCart-based system
Display19-inch high-resolution LCD monitor

Related product manuals