GE
D
IRECTION 5535208-100, REV. 2 LOGIQ E9 SERVICE MANUAL
8 - 54 Section 8-5 - Replacing Covers and Bumpers
8-5-11 Rear Cover replacement
NOTE: When replacing Rear Cover Assembly for a LOGIQ E9 used in veterinary environment, the
LOGIQ E9 Spare Parts Handling for Veterinary Systems MUST BE reinstalled before the
LOGIQ E9 returned for use. See: 8-5-8-5 "Label Placement for the LOGIQ E9 used in a
Veterinary Environment" on page 8-51 and Section 9-20 "Product Labels on LOGIQ E9
consoles used in a veterinary environment" on page 9-113.
If the V Nav On-Board Stand Option is present on the LOGIQ E9, the Option must be removed to
remove the Rear Cover. See:
8-12-5 "Assembling or replacing the On-Board V Nav Stand" on page 8-310.
8-5-11-1 Manpower
One person, 15 minutes.
8-5-11-2 Tools
For tools needed, refer to: 8-2-5 "Tools needed for servicing the LOGIQ E9" on page 8-5.
8-5-11-3 Preparations
When preparing for the replacement, you must perform the following steps:
1.) Power down the system
2.) Disconnect the mains power cable from the wall outlet.
3.) Disconnect all Probes and External I/O Cabling.
4.) Remove the Rear Bumper.
5.) Remove both Side Covers.
6.) Remove Filter Cover at rear.
7.) Remove Filter.
Follow these links if you need more information:
• 4-2-3 "Power shut down" on page 4-7.
• 8-5-2 "Side Covers replacement" on page 8-30.
• 8-5-9 "Filter Cover replacement" on page 8-52.
• 8-7-2 "Rear Filter and “handle type“ Bottom Filter replacement / cleaning" on page 8-176.