EasyManua.ls Logo

GE LOGIQ E9

GE LOGIQ E9
808 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE
D
IRECTION 5535208-100, REV. 2 LOGIQ E9 SERVICE MANUAL
Chapter 1 Introduction 1 - 25
1-9-3 Electrostatic discharge (ESD) prevention
WARNINGWARNING
DO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO TAKING THE
NECESSARY ESD PRECAUTIONS:
1. ALWAYS CONNECT YOURSELF, VIA AN ARM-WRIST STRAP, TO THE ADVISED ESD
CONNECTION POINT LOCATED ON THE REAR OF THE SCANNER (TO THE RIGHT OF
THE POWER CONNECTOR).
2. FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE
EQUIPMENT.
WARNINGWARNING
Risk of electrical shock, system must be turned off. Avoid all contact with electrical
contacts, conductors and components. Always use non-conductive handles designed
for the removal and replacement of ESD sensitive parts. All parts that have the potential
for storing energy must be discharged or isolated before making contact.
If the Shear Wave Option is present, MAKE SURE the LEDs on the Capacitor Pack are
OFF before disconnecting the Capacitor Pack Cables.

Table of Contents

Other manuals for GE LOGIQ E9

Related product manuals