IRECTION 5535208-100, REV. 2 LOGIQ E9 SERVICE MANUAL
8 - 186 Section 8-7 - Main Console parts replacement
8-7-3-3 Calibration and adjustments
No calibrations or adjustments are needed after this part replacement.
8-7-3-4 Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Connect cables and Probes you removed earlier
2.) Power up the system to verify that it operates as intended.
8-7-3-5 Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
If all are successful, include the following debrief script: LOGIQ E9 Service Manual, Direction
5535208-100, Rev. 2. Equipment passed all required checks and is ready for use
Table 8-111 Filter replacement Functional Checks
See:
Section
Functional Check Debrief Script
4-2-2 Power ON/Boot Up
LOGIQ E9 Service Manual, Direction
5535208-100, Rev. 2. Equipment passed all
required checks and is ready for use.
4-2-3 Power shut down
4-3-14 Probe/Connectors Checks
4-3-5 B-Mode Checks
4-3-7 System CFM and PWD Checks
4-3-18 Mechanical Function Checks