IRECTION 5535208-100, REV. 2 LOGIQ E9 SERVICE MANUAL
8 - 316 Section 8-12 - V Nav Roll Stand and/or On-Board Stand Installation and/or Replacement
8-12-7 Routine Maintenance
1.) Periodically inspect all fasteners associated with the mounting system. Tighten them as necessary
for optimal operation and safety.
NOTE: DO NOT overtighten any fasteners.
8-12-8 Cleaning the Mounting Assembly
1.) The mounting assembly may be cleaned with most mild, non-abrasive solutions commonly used in
the hospital environment (e.g., diluted bleach, ammonia, or alcohol solutions).
2.) The surface finish will be damaged permanently by strong chemicals and solvents such as acetone
and trichloroethylene.
3.) Do not use steel wool or other abrasive material to clean the mounting assembly.
4.) The damage caused by the use of unapproved substances or processes will not be covered by
warranty. It is recommended to testing any cleaning solution on a small area of the mounting
assembly, that is not visible, to verify compatibility.
5.) Never submerge the roll stand nor allow liquids to enter the mounting assemblies. Wipe any
cleaning agents off of the mounting assembly immediately, using a water-dampened cloth. Dry all
mounting assemblies thoroughly after cleaning.
CAUTION
THE COMPANY MAKES NO CLAIMS REGARDING THE EFFICACY OF THE LISTED CHEMICALS
OR PROCESSES AS A MEANS FOR CONTROLLING INFECTION. CONSULT YOUR HOSPITAL’S
INFECTION CONTROL OFFICER OR EPIDEMIOLOGIST. TO CLEAN OR STERILIZE MOUNTED
INSTRUMENTS OR ACCESSORY EQUIPMENT, REFER TO THE SPECIFIC INSTRUCTIONS
DELIVERED WITH THOSE PRODUCTS.