IRECTION 5535208-100, REV. 2 LOGIQ E9 SERVICE MANUAL
Chapter 10 Care & Maintenance 10 - 17
10-6-6 Data sheet for enclosure/chassis leakage current
Table 10-13 on page 10-17 shows a typical format for recording the enclosure/chassis leakage current.
Measurements should be recorded from multiple locations for each set of test conditions.
The actual location of the test probe may vary by Ultrasound system.
NOTE: Values in italics font are given as examples only. Record all data in the Electrical safety tests
log.
NOTE: Values in italics font are given as examples only.
Table 10-13 Typical data format for recording enclosure/chassis leakage
Unit under test___________________________ Date of test:_____________________
Test Conditions Measurement/Test Point Location
System
Power Grounding/PE
Rear
Panel
Lower
Frame
Probe
Connector
Main
Handle
Off Closed
Off Open
On Closed
On Open