10 - 28 Section 10-7 - When There’s Too Much Leakage Current...
10-7-4-1 New Unit
If the leakage current measurement tests fail on a new Ultrasound system and if situation can not be
corrected, submit a Safety Failure Report to document the Ultrasound system problem. Remove
Ultrasound system from operation
10-7-5 ECG Fails
Inspect cables for damage or poor connections.
10-7-6 Ultrasound Equipment Quality Check (EQC and IQC)
Download and use the latest version of these forms. They can be retrieved from MyWorkshop.
• EQC -- Refer to DOC0929340 in MyWorkshop