IRECTION 5535208-100, REV. 2 LOGIQ E9 SERVICE MANUAL
Chapter 3 LOGIQ E9 Setup 3 - 31
3-6-9 Complete power down (cont’d)
3-6-9-1 Sleep Mode
Only available in LOGIQ E9 running R2.x.x and R3.x.x software.
Use Sleep Mode when you do a portable exam in order to reduce the time to start up the system. When
you use Sleep Mode, it only takes 90 seconds to start up the system versus 2-3 minutes. To activate
Sleep Mode, press the ON/OFF button and select Sleep.
Before returning a system to the customer, perform the functional checks. See: Section 4-3 "Functional
checks" on page 4-31.
NOTICE
For optimum system operation, it is recommended that a full shutdown of the system is performed at
least once every 24-hour period. If you shut down the system at the end of the day, no other action is
needed.

Table of Contents

Other manuals for GE LOGIQ E9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ E9 and is the answer not in the manual?

GE LOGIQ E9 Specifications

General IconGeneral
TypeUltrasound System
3D/4D ImagingYes
Transducer Ports4 active ports
ConnectivityDICOM, USB, Ethernet
ApplicationsAbdominal, Vascular, Musculoskeletal, Small Parts, Urology, Pediatric, Cardiac, Obstetrics/Gynecology
TransducersLinear, Convex, Phased Array, Volume
Doppler ModesColor Doppler, Power Doppler, Pulsed Wave Doppler, Continuous Wave Doppler
ElastographyYes
Advanced FeaturesB-Flow, Volume Navigation
PortabilityCart-based system
Display19-inch high-resolution LCD monitor

Related product manuals