Cleaning and Disinfection
Sodium salts
Inspect the Device
Carefully inspect devices between uses to verify proper function.
Evidence of damage and wear on a device may include but is not limited to
discoloration, excessive scratches, wear, and cracks. Improperly functioning devices,
damaged, and excessively worn devices should not be used.
Care at the Point of Use
Clean instruments as soon as possible after use.
Soiled devices must be separated from non-contaminated devices to avoid
contamination of personnel or surroundings.
Preparation for Cleaning
For multi-piece or complex instruments, please refer to their disassembly instructions.
The disassembly instructions are available in the MAC VU360
Resting ECG Analysis
System Service Manual.
Contact your local GE Healthcare service representative for further information.
For instruments produced by another manufacturer, reference the manufacturer’s
instructions for use.
Visual Inspection, Cleaning and Disinfection Frequency
The table indicates the frequency of visual inspection, cleaning, and disinfection
procedures.
Component Visual Inspection Cleaning Disinfection
Device and Trolley Daily, preferably before
the equipment’s first
use each day
Monthly, or more
frequently, as needed
Follow the same
frequency as cleaning.
Disinfection must
be performed after
cleaning.
Acquisition Module Daily After use on each
patient and prior to use
on subsequent patient.
After use on each
patient and prior to use
on subsequent patient.
Disinfection must
be performed after
cleaning.
Leadwires Refer to the supplier's instructions for leadwire cleaning and disinfection.
114 MAC VU360
Resting ECG Analysis System 2088531-370-2

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE MAC VU360 and is the answer not in the manual?

GE MAC VU360 Specifications

General IconGeneral
Device TypeElectrocardiograph
Power100-240 V AC, 50/60 Hz
Battery TypeLithium-ion
ConnectivityUSB
Data StorageInternal memory, USB drive
Sampling Rate1000 samples per second
Battery LifeUp to 4 hours

Related product manuals