Recording an ECG or Rhythm
If the device is configured to print the barcode of the patient ID in the patient reports,
the printed patient report includes the barcode. The barcode can be used to perform
a query by patient ID in the MUSE system.
If a printer error occurs, the progress message is replaced by the related printer error.
The printing restarts automatically after the error is resolved. For more information on
printer errors, see Printing Errors on page 123.
To stop printing a patient report, do one of the following:
• Press the Stop key on the keyboard:
• Press the Stop button on the Acquisition Module:
All pending print jobs are cancelled.
Record a Rhythm
Ensure that sufficient paper is available in the paper tray to print a rhythm report.
A rhythm report can be stored in digital form in the Files list or printed on paper,
depending on how the device has been configured for your site. A digital rhythm
report cannot be transmitted to a configured automatic destination.
1. Start a test for a new patient. See Start a Test for a New Patient on page 31.
2. Change the lead set or format, gain, speed, or filter, if required. See Change Lead
Sets and Lead Formats on page 44.
3. Start the rhythm for the patient in one of the following ways:
• Select the Start Rhythm icon on the Acquisition screen:
• Press the Rhythm key on the keyboard:
• Press the Rhythm button on the Acquisition Module:
2088531-370-2 MAC VU360
™
Resting ECG Analysis System 57