Technical Specifications
Item Description
Downsampled ECG waveform Bandwidth: 0.04 to 150 Hz
Sample rate: 2 ksps
Resolution: 1.22 µV
Analyzed ECG waveform Bandwidth: 0.04, 0.56 ZPD to 150 Hz
Sample rate: 500 and 1000 sps
Resolution: 4.88 µV
Additional report filters 20 Hz, 40 Hz, 100 Hz, or 150 Hz
Channels Up to 15 channels
Environmental Specifications
Table 61: Environmental Specifications
Item Description
Operating Conditions
Temperature 10ºC to 40ºC (50° to 104° F)
Relative Humidity (RH) 20% to 95% (non-condensing)
Atmospheric Pressure 70 to 106kPa
Transport/Storage Conditions
Temperature -20ºC to +60ºC (–4° to 140° F)
Relative Humidity (RH) 15% to 95% (non-condensing)
Atmospheric Pressure 50kPa to 106kPa
Safety Specifications
Table 62: Safety Specifications
Item Description
Certification marks
The medical device has a lifetime of 7 years with respect to the
Council Directive 93/42/EEC essential requirement #4.
164 MAC VU360
Resting ECG Analysis System 2088531-370-2

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE MAC VU360 and is the answer not in the manual?

GE MAC VU360 Specifications

General IconGeneral
Device TypeElectrocardiograph
Power100-240 V AC, 50/60 Hz
Battery TypeLithium-ion
ConnectivityUSB
Data StorageInternal memory, USB drive
Sampling Rate1000 samples per second
Battery LifeUp to 4 hours