Recording an ECG or Rhythm
Record a Full Disclosure ECG
Before you begin this procedure, make sure:
You purchase and enable the Full Disclosure option in the Options Manager.
You enable the Full Disclosure option in the Setting screen.
1. Start a new patient test.
2. To view a Full Disclosure ECG, click the Full Disclosure tab on the Acquisition
screen.
One lead of the Full Disclosure ECG displays. The Full Disclosure shows the
waveform from the left to right side of the screen.
The Full Disclosure ECG records for a maximum of 5 minutes. The recording
stops after the 5 minutes is complete.
The Full Disclosure screen displays 10 lines of ECG data and each line is of 30
seconds.
A notification message displays on the Acquisition screen after the Full
Disclosure ECG records for 5 minutes.
The Full Disclosure screen displays the previous 5 minutes of Full Disclosure
ECG data.
If you disconnect the acquisition module, the ECG acquisition stops in 10
seconds.
3. To change the lead, click Select Lead.
All the configured leads display. Select the single lead you want to display on the
screen and on the printed FD reports. If you connect the leads to a patient after
being fully disconnected for at least 30 seconds, all the data will be cleared from
the display.
You can record a 15-lead ECG from the Full Disclosure waveform. To do this,
select Live screen and change the mode to Adult 15 Lead.
The ECG recording restarts and the selected lead is applied to the Full Disclosure
waveform. Clears all the previous recorded data.
4. To restart the Full Disclosure ECG, click Restart. Deletes all the current waveform
data.
The message unsaved data will be lost. Confirm to proceed displays.
5. To record a 10 seconds ECG when in the Full Disclosure screen, do the steps that
follow:
68 MAC VU360
Resting ECG Analysis System 2088531-370-2

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE MAC VU360 and is the answer not in the manual?

GE MAC VU360 Specifications

General IconGeneral
Device TypeElectrocardiograph
Power100-240 V AC, 50/60 Hz
Battery TypeLithium-ion
ConnectivityUSB
Data StorageInternal memory, USB drive
Sampling Rate1000 samples per second
Battery LifeUp to 4 hours