Cleaning and Disinfection
See the Diagnostic Cardiology Supplies and Accessories Guide, 2102946-001.
Storage
Follow the storage instructions:
Store the device, trolley, and acquisition module in a clean and dry, well-ventilated
area protected from dust, moisture, insects, vermin, and extremes of temperature
and humidity.
Use below guidelines when storing acquisition modules:
Hang the device using a holder, if available.
If cables and leadwires are attached, route the additional length cables and
store them in the basket behind the device.
Other Cleaning and Disinfection Agents
Super Sani-Cloth
®
Wipes is the recommended cleaning and disinfection solution,
which has been validated on the device. However, below products are compatible
with the device and may be used for cleaning and disinfection. It should be noted that
these products have not been validated and their efficacy cannot be guaranteed.
PDI Easy Screen Cleaning
®
PDI Super Sani-Cloth
®
PDI Sani-Cloth
®
Bleach
Clinell Sporicidal Wipes
PDI Sani-Cloth
®
AF3
PDI Sani-Cloth
®
Plus
PDI Sani-Cloth
®
HB
Clorox Healthcare
®
Hydrogen Peroxide Cleaner Disinfectant Wipes
Oxivir
®
Tb Wipes
Clinell Universal Range
Cleanisept Wipes
Mikrozid Sensitive Wipes
Caviwipes
Phenol 2% (v/v)
Ethanol (ethyl alcohol) 96% (v/v)
Hydrogen Peroxide 20% (v/v)
Sodium Hypochlorite (NaOCl) 5% solution)
Isopropyl alcohol 70% (m/m)
2088531-370-2 MAC VU360