Troubleshooting
Error Message Error Condition How to troubleshoot:
Cannot perform action while
printing
You have tried to perform
below actions while printing a
patient report:
Start a test for a new
patient
Delete the patient report
Change speed, gain, or filter
Access User Menu
Access Orders, Files, or
Queue lists
Power off, lock, log out,
standby or privacy mode
Perform only the allowed
actions.
Table 32: Common Printer Errors
Error Message Error Condition How to troubleshoot:
Low Battery. Printer is
disabled. Connect power cord.
Low battery Connect power cord.
High printer temperature.
Printer disabled. Contact
Service.
High printer temperature Restart the system to check
if the error is resolved. If the
error persists, contact your GE
Healthcare Service Support
representative.
Printer error. Restart system.
Contact Service.
Unknown error or hardware
failure in printer
Restart the system to check
if the error is resolved. If the
error persists, contact your GE
Healthcare Service Support
representative.
Incompatible firmware. Incompatible printer firmware Contact your GE Healthcare
Service Support representative
to upgrade printer firmware.
Acquisition error. Fix and retry
printing.
Printing stopped due to
acquisition error
Rectify the error with the
acquisition module and retry
printing.
Printer error. Retry. If issue
persists, restart system.
Unknown error in printer Retry printing. If the error
persists, restart the system.
Printer is recovering. Please
wait ...
Printer recovery error Wait for the printer to recover.
If the error persists, restart the
system.
2088531-370-2 MAC VU360
Resting ECG Analysis System 125

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE MAC VU360 and is the answer not in the manual?

GE MAC VU360 Specifications

General IconGeneral
Device TypeElectrocardiograph
Power100-240 V AC, 50/60 Hz
Battery TypeLithium-ion
ConnectivityUSB
Data StorageInternal memory, USB drive
Sampling Rate1000 samples per second
Battery LifeUp to 4 hours