Patient Information Fields
Table 51: Patient Information text boxs
Name Description Length Accepted Values
Standard Patient ID:
1 to 16
characters
A to Z
a to z
0 to 9
All characters are supported.
Danish Patient ID:
10
characters
Valid values: '0 to 9' and '-'.
The PID must be in the format
ddmmyy-exxg or ddmmyyexxg,
where:
dd = patient day of birth
mm = patient month of birth
yy = patient year of birth
e = patient year of birth is
calculated as follows:
e = 0, 1, 2, 3 or 4, if the
patient year of birth is
between 1900 to 1999.
e = 5, 6, 7, 8 or 9, if the
patient year of birth is
between 2000 to existent
year.
xx = patient place of birth
g = patient gender
male = odd number
female = even number
Patient ID Identification number given to the
patient.
The Patient ID (PID) can be
configured by the administrator
for a specified country
requirement (for example,
Denmark, Sweden, or Norway),
customize text box names, and
add leading zeroes for specified
character lengths.
If the Patient ID does not agree
with the configuration, an error
message opens by the Patient ID
text box.
If the Patient ID agrees with the
configuration, the patient Date of
Birth and Gender text boxes are
automatically updated.
Swedish Patient ID:
2088531-370-2 MAC VU360
Resting ECG Analysis System 151

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE MAC VU360 and is the answer not in the manual?

GE MAC VU360 Specifications

General IconGeneral
Device TypeElectrocardiograph
Power100-240 V AC, 50/60 Hz
Battery TypeLithium-ion
ConnectivityUSB
Data StorageInternal memory, USB drive
Sampling Rate1000 samples per second
Battery LifeUp to 4 hours

Related product manuals