August 2003
10-6 Specifications
Pressure Measurement
Note: An optional pressure sensor is available for TF Series and
Moisture Image Series moisture probes.
Ranges: 30 to 300 psig
50 to 500 psig
100 to 1,000 psig
300 to 3,000 psig
500 to 5,000 psig
Accuracy: ±0.5% at full scale
Proof Pressure: 3 times span of available range up to
maximum 7,500 psig
Sample System
Components: Inlet needle valve, a built-in coalescing
filter and sample cell with bypass needle
valve and venting tube, pressure gage
(various ranges), outlet needle valve
with venting tube.
Wetted Parts: 316 SS
Operating Pressure: Configurations available for 300, 500,
1,000, 2,000 or 3,000 psig (dependent on
pressure gage)
Maximum Pressure
Rating: Standard: 3,000 psig
Optional: 5,000 psig
Options: Inlet pressure regulator, 0 to 500-psig
outlet.
Armored flowmeter, 1.3 to 13 SCFH
3-m (10-ft) flexible, stainless steel-
braided, Teflon-lined hose with 1/8-in.
tube connections. Not recommended for
dew point temperatures below -75°C
(-103°F).