August 2003
3-6 Programming Site Data
Entering Calibration Data
Along with basic probe parameters, the PM880 requires 
calibration data for its probes. You will need to enter this data at 
initial setup, if you change or replace a moisture probe, or if you 
have sent the probes back to the factory for calibration. 
IMPORTANT: Be sure you have the Calibration Data Sheets 
that are supplied with each GE Panametrics 
probe. Each Calibration Data Sheet consists of a 
list of data points that you will need to enter or 
verify, and lists its corresponding probe serial 
number. Calibration Data Sheets are usually 
packed inside the probe cases.
You only need to enter calibration data for M and TF Series 
probes. The Moisture Image Series probe stores calibration data 
in its electronics module, and uploads it into the PM880 memory 
when needed. It is unnecessary to enter calibration data for the 
Moisture Image Series probe, unless you send it back to GE 
Panametrics for calibration without its electronics module. In this 
case, enter the calibration data for the recalibrated probe as 
described on the following pages.
From the Program Menu, press the [
T] key once to reach the 
Calibration option, and press 
[ENTER]. The screen appears 
similar to Figure 3-3 on the next page.