Programming Site Data 3-11
August  2003
Entering Saturation Constants
The PM880 requires a saturation constant in order to calculate 
ppm
w
 in nonaqueous liquids. (If you do not know the saturation 
constants of the process liquid sample, contact GE Panametrics.) 
To enter a saturation constant, you must enter 1 to 8 data points to 
represent a curve of Cs (constants) versus temperature. Complete 
the following steps:
1. To enter the Saturation window, use the arrow keys to reach 
the Saturation tab and press 
[ENTER]. The screen appears 
similar to Figure 3-5 below.
Figure 3-5:  The Saturation Window in the Calibration 
Option
2. Press the [
T] key to enter the first temperature value (here 
shown as C) in the saturation table.
3. Press 
[ENTER] to open the temperature text box. Use the 
numeric keys to enter the desired value, and press 
[ENTER] to 
confirm the entry. 
4. Press the [
X] key to move to the Cs text box, and press 
[ENTER]. Again, use the numeric keys to enter the desired 
value, and press 
[ENTER] to confirm the entry.
Calibration - [0000000]
Delete
Exit
Hygro Pressure Saturation Probe ID
## Temp C Cs
1
2
3
4
5
- - - -
- - - -
- - - -
- - - -
- - - -
- - - -
- - - -
- - - -
- - - -
- - - -
Insert