EasyManua.ls Logo

GE PM880 - Page 206

GE PM880
249 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4/9/98
A-24 Application of the Hygrometer (900-901D1)
Table A-3: Maximum Liquid Flow Rates
Based on the physical characteristics of benzene at a
temperature of 77°F, the following flow rates will produce
the maximum allowable fluid linear velocity of 10 cm/sec
in the corresponding pipe sizes.
Inside Pipe
Diameter (in.)
Flow Rate
(gal/hr)
Flow Rate
(l/hr)
0.25 3 11
0.50 12 46
0.75 27 103
1.0 48 182
2.0 193 730
3.0 434 1,642
4.0 771 2,919
5.0 1,205 4,561
6.0 1,735 6,567
7.0 2,361 8,939
8.0 3,084 11,675
9.0 3,903 14,776
10.0 4,819 18,243
11.0 5,831 22,074
12.0 6,939 26,269

Table of Contents

Other manuals for GE PM880

Related product manuals