Application of the Hygrometer (900-901D1) A-37
4/9/98
C. Additional Notes for Liquid Applications
In addition to the topics already discussed, the following general 
application notes pertain to the use of GE Panametrics moisture 
probes in liquid applications:
1. All M Series Aluminum Oxide Moisture Sensors can be used 
in either the gas phase or the liquid phase. However, for the 
detection of trace amounts of water in conductive liquids (for 
which an empirical calibration is required), the M2 Sensor is 
recommended. Since a background signal is caused by the 
conductivity of the liquid between the sensor lead wires, use 
of the M2 Sensor (which has the shortest lead wires) will 
result in the best sensitivity.
2. The calibration data supplied with GE Panametrics Moisture 
Probes is applicable to both liquid phase (for those liquids in 
which a Henry’s Law analysis is applicable) and gas phase 
applications.
3. As indicated in Table A-3, the flow rate of the liquid is limited 
to a maximum of 10 cm/sec.
4. Possible probe malfunctions and their remedies are discussed 
in the Troubleshooting chapter of this manual.