Programming Site Data 3-17
August  2003
Entering the System Configuration (cont.)
2. The next prompt, K × PPMv, asks for a constant by which the 
meter can multiply the ppm
v 
value for special purposes. (The 
default multiplier is 1.)
a. Press 
[ENTER] to open the text box. 
b. Use the numeric keys to enter the desired interval (in 
minutes).
c. Press 
[ENTER] to confirm the entry.
3. The PM880 now asks for the response type. Three choices are 
available: NORM (normal), CER (computer enhanced 
response), and DEF (derivative fast response).
a. Press 
[ENTER] to open the drop-down menu. 
b. Use the  [
T] and [S] keys to move to the desired 
response. 
c. Press 
[ENTER] to confirm the entry.
If you select NORM or CER, you have finished entering 
parameters in this option. But if you select DEF, you must also 
enter time constants for the sampling system (T System) and for 
the moisture sensor (T Sensor).
4. To enter the T System constant:
a. Press 
[ENTER] to open the text box. 
b. Use the numeric keys to enter the desired interval (in 
minutes).
c. Press 
[ENTER] to confirm the entry.
5. Repeat step 4 to enter the T Sensor constant.
• To confirm the data and return to Operate Mode, press [F3] 
(OK). The PM880 returns to Operate Mode.
• To leave the window without confirming the data, press [F2] 
(Cancel) or the 
[ESC] key.  The PM880 returns to Operate 
Mode.