EasyManua.ls Logo

GE Ultrasonic USM Go - Protection

GE Ultrasonic USM Go
190 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
USM Go Operators Manual 141
A.8 Protection
Damp Heat & Humidity (Storage): 10 Cycles: 10 hr at +60°C down to +30°C, 10 hr at +30°C up to +60°C,
Transition within 2 hr (507.4)
Temperature Shock (Storage): 3 Cycles: 4 hr at -20°C up to +60°C, 4 hr at +60°C,
Transition within 5 minutes (503.4 Procedure II)
Vibration: General Exposure: 1 hr each axis, 514.5-5 Procedure I, Annex C, Figure 6
Shock: 6 Cycles each axis, 15 g, 11 ms, half-sine (516.5 Procedure I)
Loose Cargo (in Shipping Container): 514.5 Procedure II
Transit Drop (Packaged for Shipment): 26 Drops, 516.5 Procedure IV
Dust Proof/Dripping Water Proof: As per the IEC 529 specification for IP67 classification
Operating Temperature Range: 0 to 55°C
Storage Temperature Range: -20 to +60°C, 24 hr with battery