Chapter 2. Instrument Setup
USM Go Operators Manual 53
2.5.2 Setting the Display Delay
The display delay function shifts the displayed A-Scan to the left or the right in the viewing window. To set the display delay:
1. In the ACQUIRE menu, activate the RANGE submenu using the joystick ( ).
2. Use the joystick ( ) to select the function titled DISPLAY DELAY.
3. Press the function toggle or move the joystick ( ) to change the display delay. As you change the value, you should see the displayed
echoes shift to the left or the right.
Note: Pressing both buttons on the Function Toggle simultaneously resets the DELAY to the default value of 0.
4. After completing your selection, move the joystick up or down ( ) to navigate away from this function.