EasyManua.ls Logo

GE Ultrasonic USM Go - 2.5 Adjusting the A-Scan; 2.5.1 Setting the A-Scan Range; 3.1.3 Setting Gate Alarms and Outputs

GE Ultrasonic USM Go
190 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 2. Instrument Setup
USM Go Operators Manual 51
2.5 Adjusting the A-Scan
To configure the USM Go A-Scan, follow the instructions in this section.
2.5.1 Setting the A-Scan Range
Calibration of the USM Go requires the use of two calibrated standards. These standards must be of different thicknesses, and they must be
made of the same material as the test piece. Prior to calibrating the instrument/probe combination, the the A-Scan display range, which is the
material thickness value represented by the full horizontal width of the screen, is normally set to a value equal to or slightly larger than the
thicker calibrated standard (see Figure 13 below).
Figure 13: Effects of the A-Scan Range Adjustment

Table of Contents

Related product manuals