Appendix C. EN 12668 Specifications
USM Go Operators Manual 153
Paragraph Parameter LSL Typ. USL Units Test Conditions
9.4.2 Pulse Reverberation 0 4 nSec Damping = 50, Voltage = Low, Energy = Low, RepRate = 15
0 4 nSec Damping = 50, Voltage = Low, Energy = Low, RepRate = 2000
0 4 nSec Damping = 1000, Voltage = Low, Energy = Low, RepRate = 15
0 4 nSec Damping = 1000, Voltage = Low, Energy = Low, RepRate = 2000
0 4 nSec Damping = 50, Voltage = Low, Energy = High, RepRate = 15
0 4 nSec Damping = 50, Voltage = Low, Energy = High, RepRate = 2000
0 4 nSec Damping =1000, Voltage = Low, Energy = High, RepRate =15
0 4 nSec Damping =1000, Voltage = Low, Energy = High, RepRate =2000
0 4 nSec Damping =50, Voltage = High, Energy = Low, RepRate =15
0 4 nSec Damping =50, Voltage = High, Energy = Low, RepRate =2000
0 4 nSec Damping =1000, Voltage = High, Energy = Low, RepRate =15
0 4 nSec Damping =1000, Voltage = High, Energy = Low, RepRate =2000
0 4 nSec Damping =50, Voltage = High, Energy = High, RepRate =15
Table 2: EN 12668-1:2000 Specifications (cont.)