Appendix C. EN 12668 Specifications
USM Go Operator’s Manual 157
Paragraph Parameter LSL Typ. USL Units Test Conditions
9.5.5 Linearity of Vertical Display -2 0.6 2 %FSH Low Pass Selected
-2 0.6 2 %FSH 4 to 5 MHz Selected
-2 -0.9 2 %FSH 10 MHz Selected
-2 -1.25 2 %FSH 13 MHz Selected
-2 0.18 2 %FSH BroadBand Selected
8.7.2 Linearity of Timebase - 0.03 0.5 %FSW
8.7.3 Digitization Sampling Error -5 -4 5 %FSH
Dimensions 17.1 CM Height
28.2 CM Width
15.9 CM Depth
3.8 KG Weight (With Battery)
Table 2: EN 12668-1:2000 Specifications (cont.)