Chapter 2. Instrument Setup
48 USM Go Operators Manual
2.4.8 Selecting a Rectification Mode (cont.)
Positive Half Rectification means that only the upper (i.e. positive) half of the RF signal is displayed.
Negative Half Rectification means that only the bottom (i.e. negative) half of the RF signal is displayed (see Figure 12 on page 47). Note
that although only the negative half of the RF signal is displayed, it is shown in the same orientation as a positive component to simplify
viewing.
Full-Wave Rectification combines the positive- and negative-rectified signals together, and displays both of them in a positive
orientation (see Figure 12 on page 47).
Use the procedure on the next page to select a rectification mode.