Diagnostics/Troubleshooting
7-48 System view Versana Balance – Basic Service Manual
5808768-100 English Rev.9
7-6-3-6-1e-PAT(continued)
6. If you want to review all Reports to ensure the probe is fully
functional and has no damaged elements, go to In the Insite
ExC > Service Desktop > Diags > Diags History to
check the probe diagnostic results.
7. All the probe diagnostics you have run will be displayed in
Runs field. Select one and press Get History button, the
corresponding reports will be shown in the Diagnostic
Results History field.