Calibration Zoom | 177
Mechanical cali-
bration
The following instrument parts can be calibrated mechanically:
Circular level on the instrument and tribrach.
•Laser plummet.
Screws on the tripod.
)
During the manufacturing process, the instrument errors are carefully deter-
mined and set to zero. As mentioned, these errors can change and it is highly
recommended to redetermine them in the following situations:
Before the instrument is used for the first time.
Before every high precision survey.
After rough or long periods of transport.
After long periods of work or storage.
If the temperature difference between current environment and the
temperature at the last calibration is more than 10°C (18°F).