WindMaster and WindMaster Pro 1561-PS-0001. Issue 11 February 2017
________________________________________________________________________________________________________________________
52
Ax – Select SOS/Sonic Temp output
1 Neither
2 SOS
3 Sonic Temp
4 Both
The SOS and/or Sonic Temp are displayed after the UNITS and before the status byte, the SOS is always in m/s and the
Sonic Temp is always in deg C.
Bx – Set Baud rate
1 2400
2 4800
3 9600
4 19200
5 38400
6 57600
If a request is sent to change the Baud rate, before it changes it must be confirmed by entering B ENTER at the new
Baud rate.
E.g. If set to B3 (9600 baud), to change to B5 (38400 baud), enter B 5 ENTER , change host terminal to 38400
baud, and confirm by entering B ENTER.
NOTE: a random echo may be generated after the B5 confirmation
Certain combinations of output rate, baud rate and message type may be unsupported.
Cx – Analogue output polar direction wrapping
1 540 degree wraparound on analogue output
2 360 degree wraparound on analogue output
Dx – Diagnostic and configuration information
1 Request serial number
2 Request SW version
3 Request current configuration
4 -
5 -
6 -
7 Report DAC channel configuration
Examples of Diagnostic information are as follows:
Request Serial Number
D1
W154503 or Y154503 (see Para 5.3 for explanation).
Request SW Version
D2
2329-700