2-3
CONNECT CALIBRATION GAS AND POWER SUPPLY
The concentrations, illustrated in Fig. 2-2, should be consistent
with Auto Bump / Cal Station installed software conguration
window, as illustrated in Fig. 2-3.
Software
Configuration
Gas Concentr
Fig. 2-3 Software Conguration
After comparing gas values between Cylinder Gas Label and
Software Conguration, when consistent, proceed to next
paragraph.
If inconsistent, refer to Chapter 3 to update software
conguration then download updated conguration to Auto
Bump / Calibration Station.