3-6
AUTO BUMP / CAL STATION - HANDBOOK
The test times can be set from 15 to 60 seconds, as illustrated
in Fig. 3-7.
To edit test time, position cursor on value required then left
click mouse button to conrm.
3.1.2 Full Bump Test (Factory Default)
This feature is the factory set default option and provides the
user with the ability to perform a full bump test then, if selected
as illustrated in Fig. 3-8, follow up with an automatic calibration
of failed gas ranges, if applicable.
Fig. 3-8 Default Selection
When gas is applied, using the GMI Test Gas Cylinder (Part
No. 99146), for a pre-set 60 seconds and when the instrument
tolerance band listed below, is reached for each gas range,
the Bump / Cal Station will record a ‘Pass’.
The instrument default gas range tolerances are as follows:
LEL hydrocarbons: 45% to 62%
Oxygen (O
2
): 18.5% to 17.5%
Carbon Monoxide (CO): 450ppm to 550ppm
Hydrogen Sulphide (H
2
S): 45ppm to 55ppm