2-2
AUTO BUMP / CAL STATION - HANDBOOK
2.1.1 Compare Gas Cylinder Label with Software
Conguration
First of all, make sure that the cylinder is not older than the
‘use by’ date on the label illustrated in Fig. 2-2.
Make sure that the cylinder contains enough gas to complete
the bump test / calibration. The pressure gauge on the
Regulator Valve, after tting, gives a clear indication of this.
When applying gas to the instrument via the Auto Bump /
Cal Station, it is important that the gas concentrations listed
on the cylinder label are consistent with the current software
conguration downloaded from the USB memory stick to the
Auto Bump / Cal Station.
As illustrated in Fig. 2-2, the cylinder gas concentrations can
be easily found on the cylinder label.
Fig. 2-2 Gas Cylinder Label