AUTO BUMP / CAL STATION - HANDBOOK
2.1.1 Compare Gas Cylinder Label with Software
Conguration
First of all, make sure that the cylinder is not older than the
‘use by’ date on the label illustrated in Fig. 2-2.
Make sure that the cylinder contains enough gas to complete
the bump test / calibration. The pressure gauge on the
Regulator Valve, after tting, gives a clear indication of this.
When applying gas to the instrument via the Auto Bump /
Cal Station, it is important that the gas concentrations listed
on the cylinder label are consistent with the current software
conguration downloaded from the USB memory stick to the
Auto Bump / Cal Station.
As illustrated in Fig. 2-2, the cylinder gas concentrations can
be easily found on the cylinder label.
Fig. 2-2 Gas Cylinder Label