ASR-2000 Programming Manual
90
<bool>|OFF|ON
This parameter is w/o function.
Return parameter
<NR2>,<NR2>,<bool>,<NR2>,<bool>,<NR1>,<NR1>,
<bool>,<NR1>,<NR1>,<bool>,<NR1>,<bool>, <bool>
Returns the common parameters in the following order:
Step time, on phase, on phase on/off, off phase, off
phase on/off, term settings, jump step number, jump
on/off, jump count, code on/off, branch1, branch1
on/off, branch2, branch2 on/off, trig out on/off.
Example1
:SEQ:CPAR 1,0,10,1,HOLD,10,1,0,1,0,0,0,0,1
Example2
:SEQ:CPAR?
>+0.1000,+0,+0,+0,+0,CONT,+1,+1,+1,+0,+0,+0,+0,+0,
+0
[:SOURce]:SEQuence:CSTep
Query
Description
Returns the currently running step number.
Query Syntax
[:SOURce]:SEQuence:CSTep?
Return parameter
<NR1>
Current step number
Example
:SEQ:CSTep?
>1
[:SOURce]:SEQuence:SPARameter
Set
Query
Description
Sets or queries the parameters for a specified step.
Syntax
[:SOURce]:SEQuence:SPARameter
{<NR2>,<NR1>|CONSt|KEEP|SWEep,<NR2>,<NR1>|C
ONSt|KEEP|SWEep,<NR2>,<NR1>|CONSt|KEEP|SWE
ep,SIN,<NR1>}
Query Syntax
[:SOURce]:SEQuence:SPARameter?
Parameter
<NR2>
ACV setting
<NR1>|CONSt|
KEEP|SWEep
ACV mode: Constant(1) | Keep(2)
| Sweep(3)
<NR2>
DCV. Not applicable. This
parameter will be ignored.
<NR1>|CONSt|
KEEP|SWEep
DCV mode: Constant(1) | Keep(2)
| Sweep(3)
<NR2>
Frequency