REMOTE CONTROL
131
Description
<value> 50-1000
I represents an infinite loop. N represents a finite
loop. The cycle number is set by parameter <value>
Example
:DELAy2:CYCLEs N,100
Sets the cycle number of Delay as 100 for CH2.
Command
:DELAy[1|2]:CYCLEs?
Function
Queries the cycle number of Delay.
Example
:DELAy2:CYCLEs?
Return to I or N, <value> for CH2 (For example:
N,100)
Command
:DELAy[1|2]: ENDState {ON|OFF|LAST}
Function
Sets the termination state of Delay.
Description
ON|OFF|LAST
ON: output enabled
OFF: output disabled
LAST: stay at the last group of output state.
Example
:DELAy2: ENDState OFF
Sets the termination state of Delay as output
disabled for CH2.
Command
: DELAy[1|2]: ENDState?
Function
Queries the termination state of Delay.
Description
ON|OFF|LAST
ON: output enabled
OFF: output disabled
LAST: stay at the last group of output state.