REMOTE CONTROL
135
Queries the currently used pattern of the
automatically generated state.
:DELAy2:STATe:GENE?
Queries the currently used pattern of the
automatically generated state for CH2.
:DELAy[1|2]:STOP
{NONE|<V|>V|=V|<C|>C|=C|<P|>P|=P}[,<value>]
Sets the stop condition of delay.
NONE: Sets stop condition type as “NONE”.
<V, >V,=V: Sets stop condition type as “<voltage”
“>voltage” “=voltage”.
<C,>C, =C: Sets stop condition type as “<current”
“>current” “=current”.
<P,>P, =P: Sets stop condition type as “<power”
“>power” “=power”.
<value> It is used to set voltage, current or power
value of stop condition, which range from 0 to the
maximum voltage/current/power value of the
active channel.
:DELAy2:STOP >V,8
Sets the stop condition of delay as >8V for CH2.
Queries the stop condition of the current delay.
:DELAy2:STOP?
Queries the current stop condition, back to NONE
or stop condition value.
:DELAy[1|2]:TIME:GENErate
{FIX|INC|DEC}[,<value0>[,<value1>]]