Command
:STATus:QUEue:ENABle?
Function
Read the error and status messages that have been
enabled.
Example
:STATus:QUEue:ENABle?
Returns the contents of the enabled error and
status messages.
Command
:STATus:QUEue:DISable <list>
Function
Specifies which messages will not be placed in the
error queue.
Description
<list>
(-440:+900): Full range error messages.
(-110): Single error message.
(-110:-222): A specific range of error
messages.
(-110:-222, -220): A specific range of
error messages and a single error
message (separated by a comma.).
Example
:STATus:QUEue:DISable (-110:-222)
The error messages in the range of -110 to -222
will not appear in the error queue.
Command
:STATus:QUEue:DISable?
Function
Reads the disabled messages.
Example
:STATus:QUEue:DISable?
Returns the disabled messages.
Command
:STATus:QUEue:CLEar
Function
Empty all the messages from the error queue.
Example
:STATus:QUEue:CLEar
Empty all the messages from the error queue.