20
Installation
3.9.2 Create new measurement configuration table
From the application’s File menu, select New and then O2 Instrument to create a new
measurement configuration table.
Note: The other two options, CO2 Instrument and Dual Channel Instrument are for future use.
A default entry is created automatically, with an ID of “0”, a Location Name and Product Name
of “Default”. None of these fields can be changed.
Use the same functionality as described for the user table, to add entries to the table. The
following information is required:
• Entry ID - 5 character numeric
• Location Name - 8 character alphanumeric
• Location Description - 32 character alphanumeric
• Product Name - 8 character alphanumeric
• Product Description - 32 character alphanumeric
• Measurement Mode - Continuous or Sample
• Sample Type - Select from a drop-down list of valid entries
• O2 Low Level Activation - Yes or No
• O2 Low Level Alarm - If activation set to yes, enter the low level value to trigger the alarm
• O2 High Level Activation - Yes or No
• O2 High Level Alarm - If activation set to yes, enter the high level value to trigger the alarm
• O2 Gas Unit Type - Select from a drop-down list of valid entries
• O2 Gas Unit - Select the display unit from a drop-down list of valid entries
• O2 Calibration Offset - Enter a calibration offset value if required
Once the table is complete, select Close, Save or Save as from the File menu to save the file
to disk.
Once a table has been saved to disk, it can be edited using the Open option from the File
menu, to open an existing table.
The table can also be printed using the Print option in the File menu.