P/N 53063-30, Manual revision: B
6-10
Cell Saving using Manual Operation
SUMMARIZING PARAMETER VARIATIONS
The following table summarize the different values for default settings used
throughout the various states of the cell saving protocol and which are
recommended for use during manual operation.
Table 6-1, Pump rate range
State 70mL Bowl Latham 125 Latham 225
FILL 125 ml/min 300 ml/min 600 ml/min
WASH 100 ml/min 300 ml/min 500 ml/min
EMPTY 100 ml/min 100 ml/min 300 ml/min
Table 6-2, Wash volume per cycle
Procedure
type
70mL Bowl Latham 125 Latham 225
Standard 300 ml 750 ml 1000 ml
Orthopedic 400 ml 1000 ml 1500 ml
Table 6-3, Maximum centrifuge speed
State 70mL Bowl Latham 125 Latham 225
FILL 7000 rpm 5650 rpm 5650 rpm
WASH 7000 rpm 5650 rpm 5650 rpm