P/N 53063-30, Manual revision: B
5-9
Cell Saving using Automatic Operation
To avoid a red cell spillage into the waste bag, the operator can:
" Press the Pause key and stop the pump.
The word “Pause” will flash on the screen display.
" Press the Pump slew keys to gradually adjust the pump speed.
The operator can adjust the pump to the desired speed while observing the
cells for separation.
To resume the process at the previous pump rate:
" Press the Pause key again.