MANUAL ADJUSTMENT PROCEDURE HM1508-2
C2017
16 CH 1, 10:1 Attenuator Input Capacitance Adaptation
– Locate and identify C-Trimmer C2017 in the shielded CH 1 section of the YP board.
– Set amplitude calibrator (square wave generator) to 5 kHz.
– Set attenuator CH 1 to 50 mV/div.
– Adjust C2017 for 2 parallel, straight lines.
correct overcompensate undercompensate
C2206 C2204 C2217 C2223 R2231 C2201 R2225
C2006 C2004 C2017 C2023 R2031 R2025