MANUAL ADJUSTMENT PROCEDURE HM1508-2
Check that no signal is applied at the BNC connectors.
Press “Adjust 2” function key to call the “Adjust Adjust 2” submenu.
Press “Calibrator” function key to call “Adjust 2 Calibrator”.
Connect a 1:1 probe with it`s tip to the PROBE ADJ. output.
Connect the 1:1 probe BNC plug with a Digital Multimeter Input.
Set Digital Multimeter to a DC measurement range suited for 0.2 V measurement with an accuracy better
than 0.1%.
– Press “0.2 V
DC
” function key.
Turn INTENS for a reading of 200 mV at the multimeter.
Press “Save” function key to store into user memory and to leave item 56.
Continue with item 57.
Check that no signal is applied at the COMPONENT TESTER sockets.
Press “Comp. Tester” function key to call “Adjust 2 Comp. Tester” menu.
Turn INTENS knob to set the Component Tester horizontal trace line on the horizontal center line of the
graticule.
Press “Save” function key.
Continue with item 58.
56 Calibrator (Software Adjustment)
57 COMP. Tester (Software Adjustment)

Table of Contents

Other manuals for Hameg CombiScope HM1508-2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hameg CombiScope HM1508-2 and is the answer not in the manual?

Hameg CombiScope HM1508-2 Specifications

General IconGeneral
BrandHameg
ModelCombiScope HM1508-2
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals